CLD 700 EL ht
Chemiluminescence NO/NO2/NOX Analyzer
The CLD 700 EL
ht two-channel nitrogen oxide analyzer is the Rolls-Royce model of the Eco Physics product
range. It can measure hot, moist sample gas directly from the source and displays the
precise NO, NOX and NO2 values.
A hot subject.
Thanks to the unique "hot
tubing", hot gas can be fed to the CLD 700 EL ht directly from the source.
Immediately downsteam of the hot tubing, the sample gas pressure is reduced to below
atmospheric pressure, ensuring that no condensation takes place within the instrument. As
a result no damage or inaccuracies occur due to the water vapour in the sample gas. In
most applications it is not necessary to use the otherwise customary sample gas cooler.
The EL ht
possesses two parallel reaction chambers, which permit simultaneous measurement and
display of NO, NOX and NO2. Pressure variations, which occur in the
sample flow, are regulated by a refined, motorized bypass system. The remaining error is
compensated for digitally. By virtue of its measurement performance and technical
capabilities, this analyzer satisfies the highest standards while leaving nothing to be
desired in terms of ease of maintenance.
Ease of operation.
The user can operate the analyzer by means
of the built-in keypad or through a PC. A simple menu structure leads the user through the
relavent procedure. The choice of options extends from setting parameters for automatic
operation to the switching on and off of individual modules. Error messages appear in full
text rather than as a code reference.
Applications:
 |
Motor vehicle and accessories
industry |
 |
Marine engine manufacturing |
 |
Generators |
 |
Incinerators |
 |
Research and development |
 |
Metal processing industry |
 |
Commerical testing and
measurement services |
This high-performance analyzer can be used
either as a stand-alone instrument or can be integrated into other systems wherever it is
necessary to measure hot gas directly at the source. For example, this measurement
technique has already been prescribed for the motor industry. In addition, regulations for
control and operating procedures have been issued for this industry (the AK protocol
agreement). We can supply appropriate firmware.
Technical Specifications:
Measuring ranges: |
0-10/0-100/0-1,000/0-10,000
ppm
(analog output adj.: 25% and 50% FS) |
Minimum detectable concentration: |
0.1 ppm |
Noise at zero point (1 sigma): |
0.05 ppm |
Zero drift: |
none |
Linearity (in range): |
± 1% FS |
Lagtime: |
5 sec |
Rise time (0-90%): |
2-32 sec
selectable (range dependent) |
Temperature range: |
10-40 °C |
Humidity tolerance: |
5-95% rel. h
(non-condensing ambient air) |
Quenching: |
for H2O:
< 0.5%/vol.-% H2O |
| for CO2: <
0.1%/vol.-% CO2 |
Sample flow rate: |
1.2 l/min |
Dry air use(for O3 generator): |
0.55 l/min |
Power required: |
660 VA |
Supply voltage: |
230 V/50 Hz,
(± 10%) optional 115 V/60 Hz (± 10%) |
Interface: |
RS 232
(standard) |
Analog output: |
selectable 1
V/10 V at 500 k, 20 mA at 600 max. |
Weight: |
40 kg |
Dimensions: |
height: 133 mm
(5 1/4"; 3 HU), with feet 158 mm (6 1/4") |
width: 483 mm
(19") |
depth: 588 mm
(23 1/8") |
Delivery includes: |
Analyzer CLD
700 EL ht, vacuum pump, bypass pump, ozone scrubber, power cables, RS 232 cable, analog
signal cable, tubing, manual |
|