CLD 700 EL
Chemiluminescence NO/NO
2/NOX Analyzer

The CLD 700 EL two-channel nitrogen oxide analyzer permits continuous simultaneous measurement and display of NO, NO2 and NOX concentrations. It can be connected directly downstream of a gas cooler and can be used alone or in parallel with other measuring instruments.

Technology that makes an impression.

cld700el1.jpg (11184 bytes)The special feature of this measuring instrument is the concept of two parallel reaction chambers. They permit simultaneous measurement of NO and NOX, allowing the instrument to display precise NO2 values without delay. The chambers are secured by means of a single screw and, consequently, easily accessible for quick cleaning when necessary.

Pressure variations in the sample, which can occur where industrial applications are involved, are removed with the assistance of a refined bypass system. The analyzer conforms not only to the highest mechanical standards but is also state-of-the-art in terms of its electronics. For example, periodic zero-point measurement and synchronization of the two channels by means of an optical positioning procedure guarantee a high level of signal stability.

Very simple operation.

cld700el2.jpg (14687 bytes)The user can operate the instrument by means of the built-in keypad or through a PC. The options extend from setting the parameters with which, for example, the automatic calibration should operate, to switching individual modules on and off. Manual operation of the CLD 700 EL, with the keyboard, is guided by simple menu-structured software.

Applications:

Gas producers
Test and calibration points
Chemical industry
Waste incineration plants
Cement producers
Paper and timber products
Glass and porcelain manufacturers

The CLD 700 EL is frequently used in installations which measure a range of waste gases and have a sample gas treatment facility. Waste incineration plants provide a typical example of this. Sample treatment in this case is essential because of the aggressive gases present. If necessary, we can also supply you with a sample gas cooler together with the instrument.

Technical Specifications

Measuring ranges: 0-10/1-100/0-1,000/1-10,000 ppm
(analog output adj.: 0-2.5/0-5/0-10/0-25 ppm etc.)
Minimum detectable concentration: 0.1 ppm
Noise at zero point (1 sigma): 0.05 ppm
Zero drift: none
Linearity (in range): ± 1% FS
Lagtime: 5 sec
Rise time (0-90%): 2-32 sec selectable (range dependent)
Temperature range: 10-40 °C
Humidity tolerance: 5-95% rel. h (non-condensing, ambient air and sample gas)
Quenching: for H2O: < 0.75%/vol.-% H2O
for CO2: < 0.15%/vol.-% CO2
Sample flow rate: 0.6 l/min
Dry air use(for O3 generator): 0.55 l/min
Power required: 500 VA (incl. membrane pump and ozone scrubber)
Supply voltage: 230 V/50 Hz, 115 V/60 Hz (± 10%)
Interface: RS 232 (standard)
Analog output: selectable 1 V/10 V/20 mA
Trigger signal output: for control of external calibration device
Weight: 35 kg
Dimensions: height: 133 mm (5 1/4"; 3 HU), with feet 158 mm (6 1/4")
width: 483 mm (19")
depth: 588 mm (23 1/8")
Delivery includes: Analyzer CLD 700 EL, membrane vacuum pump, ozone scrubber, power cables, RS-232-cable, analog signal cable, tubing, manual