Eco Physics' unique nitrogen oxide analyzer provides for the first
time continuous, exact determination of concentrations in the ppt (parts per trillion)
range. The extremely high sensitivity is guaranteed by a separate reaction chamber (to
offset cross sensitivity) and state-of-the-art signal filtering. With this hundredfold
increase in sensitivity, the CLD 770 AL ppt opens up new dimensions in monitoring air
quality, making it ideal for use in determination of required mean annual NO2
values.
The Eco Physics CLD 770 AL ppt excels not only in its high measuring
sensitivity but also in its construction. The orderly layout guarantees exactly defined
reaction conditions, while the temperature and pressure parameters, which have to be
precisely maintained, are monitored by sophisticated sensors and kept within narrow
tolerance levels. The construction, as an assembly of individual modules such as the
preliminary reaction chamber, the main reaction chamber, the ozone generator, etc. allows
fast and simple maintenance.
| Measurement
ranges: [ppb] |
0.005 to 5 } |
| 0.01 to 10 } The signal integration interval
can |
| 0.05 to 50 } be varied in increments of 1
second |
| 0.1 to 100 } (from 1 to 600 secs) for each
range. |
| 0.5 to 500 } |
| Minimum
detectable concentration: |
< 50 ppt for an integration time of 30 sec |
| < 15 ppt for an integration time of 600
sec |
| Noise (at zero, ± 2 sigma
[ppb]): |
± 50 (at signal integration time of 30 sec) |
| Signal noise: |
noise at zero (see above) or < 0.5% of
signals (at a signal integration interval of 30 sec) whichever is greater |
| Lower detectable limit: |
± 2 sigma [ppb]: < 0.100 (at a signal
integration interval of 30 sec); < 0.030 (at a signal integration interval of 600
sec) |
| Zero drift: |
No zero drift because of compensation through
the pre reaction chamber |
| Zero Point Deviation [ppb]: |
negligible |
| Linearity (in range): |
± 1% of measuring range |
| Rise time (internal): |
< 1 sec to achieve 98% of end value in
response to a step change in sample concentration alone |
| Fall time (internal): |
< 1 sec (see Rise time) |
| Response time: |
when 'measuring continuously', < 1 sec
[see Op. Instr. section 5.2.2.2 window (5)] in 'NO' and 'NO.c' modes: ^... 4 sec. |
| Temperature range: |
15-35 °C |
| Humidity tolerance: |
5-95% rel. humidity (not condensing) |
| Sample flow: |
1500 ml/min ± 100 ml/min |
| Flow rate ozone generator: |
75 ml/min |
| Power required: |
approx. 150 VA (with pump and ozone destroyer
950 VA) |
| Supply voltage: |
230 V/50 Hz (± 10%), |
| |
115 V/60 Hz (± 10%) |
| Interfaces: |
RS 232 (RS 485 optional) |
| Analog output: |
0 - 1 V / 0.2 - 1 V / 0 - 10 V / 2 - 10 V
(into min. 600 k) |
| 0 - 20 mA / 4 - 20 mA |
| (load impedance max. 600 k non-isolated) |
| Photomultiplier
temperature: |
- 100 C, selectable |
| Photomultiplier chamber
flush: |
15 mL/min (dry air) |
| Interferences [%NO2
equivalent]: |
|
NH3: |
< 1 |
SO2: |
< lower detectable limit |
water vapor: |
1% vol H2O vapor in sample gas
reduces |
| NO/NOx 0 signal by 8% |
| HNO3 (same conc. as NO2)
< 1, |
| NO3/N2O5/HO2NO2/CLONO2/PAN
< 1, HONO < 20 |
| Pump: |
Vane-type rotary vacuum pump |
| Dimensions: |
Height: 133 mm (5 1/4"=3 HE, with stand
6.22"), |
| Width: 482 mm (19") |
| Depth: 546 mm (21 1/2") |
| Weight: |
24 kg (52 lbs), Pump 24 kg |
| Delivery includes: |
Analyzer CLD 770 AL ppt, ozone destructor,
pump, filter, tubes, power cables, analog cable, RS-232 cable, manual |