5008 PROGRAMMABLE MULTI-GAS CALIBRATOR
Configurations:
Model 5008
This is the "stripped down" configuration used for
applications involving dilution of external gas cylinders for pollutants such as CO, SO2,
CH4, NO and NO2.
Model 5008 w/DET
Added to the configuration above is a pyrex reaction chamber and an
internal ozone generating subsystem controlled by an ozonator-detector feedback system. In
addition to dilution of external gas cylinders this unit has the ability to perform ozone
and Gas Phase Titration (GPT) calibrations. The detector feedback system controls ozone
output by correlating the ozonator lamps light intensity against the resulting ozone
output as measured by an external ozone standard. An internal regression analysis is
performed by the units computer to correct for any non-linearity of the data
obtained.
Model 5008 w/D&P
Added to the Model 5008 configuration is a pyrex reaction chamber and
internal ozone generating subsystem controlled by an ozonator-photometer feedback system.
As with the Model 5008 w/DET, there are the added capabilities of performing ozone and GPT
calibrations. The UV photometric feedback system controls ozone output by measuting the
actual ozone output. This system only needs to be re-calibrated as often as an ozone
primary standard.
Advantages:
 |
Multi-gas calibration via
dilution invented by Dasibi |
 |
Ozone generator controlled via
detector feedback system (Model 5008 w/DET) invented by Dasibi |
 |
Ozone generator controlled via
UV photometric feedback system (Model 5008 w/D&P) invented by Dasibi |
 |
Capable of storing up to 100
programmed gas calibration levels |
 |
Capable of storing up to 20
programmed gas calibration sequences containing up to 40 steps each |
 |
Able to accomodate up to four
external gas cylinders at any one time |
 |
Fully displayed menu-driven
software system for easy operation |
 |
Capable of performing internal
20-point calibrations on its mass flow controllers, ozone generator and UV photometer (if
present) |
 |
Use of pyrex dilution/reaction
chambers for quicker calibration gas creation |
 |
Flexibility for increasing the
number of gas mass flow controllers |
 |
Common electronics with Dasibi
Model 7001 Beta Attenuation Monitor and Model 8001 Data Acquisition System |
Technical Specifications
| Dilution
System: |
Accuracy of flow measurements: |
±
0.25% |
Repeatability of flow control: |
±
0.1% |
Linearity of flow measurements: |
±
0.25% |
Standard MFC range for dilution air: |
0-10 SLPM |
Standard MFC range for cylinder gases: |
0-100 SCCPM |
| Ozone
Generating Subsystem (Models 5008 w/DET and 5008 w/D&P) |
O3 generator output: (at 5.0 SLPM total flow; adjustable) |
0.06-1.0 ppm |
Stability: |
±
0.002 ppm @ 0.500 ppm |
Standard zero air required: |
10.0 SLPM |
| UV Photometer
Subsystem (Model 5008 w/D&P only) |
Standard range: |
0 to 1.0 ppm (capable of
measuring O3 above 1 ppm) |
Incremental sensitivity: |
0.001 ppm |
Precision: |
0.001 ppm |
Noise: |
±
0.001 ppm |
Linearity: |
Better than ±
1.0% |
Zero drift: |
Zero |
Span drift: |
Zero |
Flow rate: |
1.5 to 3.5 LPM (nominal) |
Optional analog output: |
1, 5, 10 Volts (selectable) |
Power: |
105-125 VAC, 50-60 Hz |
| 220-240 VAC, 50-60 Hz |
|