4108 U.V. Fluorescence SO2/H2S Analyzer
Configurations:
Model 4108
This is the basic configuration containing
internal zero/span valves for manually or remotely activated verification checks, on-board
zero air source for "self-zero" checks, and a particulate removal system with 5
micron Teflon filter pad.
Model 4108 w/PERM
Added to the basic configuration described
above is an internal SO2 permeating subsystem that allows "self-span"
checks. The span level created will be approximately 400 ppb, unless otherwise specified.
Model 4108 w/H2S
This combination of units consists of the
basic SO2 monitor (Model 4108) configured to be used in conjunction with the
Dasibi Model 4109 H2S to SO2 thermal oxidizer (a further description
is provided below). This allows the 4108 unit to measure SO2, H2S
and SOX.
Model 4108 w/P&H2S
This combination of units consists of the
Model 4108 w/PERM configured to be used with the Dasibi Model 4109 H2S to SO2
thermal oxidizer. This allows the 4108 w/PERM unit to measure SO2, H2S
and SOX and create a self-span check of SO2 (around 400 ppb, unless
otherwise specified).
Series 4109 H2S to SO2
Convertors
The Dasibi Models 4109 and 4109A Hydrogen
Sulfide Convertors are designed to be used in conjunction with an SO2 analyzer
to oxidize TRS compounds (i.e. H2S, MESH, DMS and DMDS) to SO2
without affecting the concentration of SO2 in the air stream. The purpose is to
allow an SO2 analyzer (such as the Model 4108) to provide a continuous stream
of high quality, non-ambiguous, Hydrogen Sulfide concentration data.
Advantages:
 |
US EPA Equivalency Designated
EQSA-1086-061 (Models 4108 & 4108 w/PERM only) |
 |
State-of-the-art non-pulsed
ultraviolet fluorescence technique specific to SO2 |
 |
First EPA-approved SO2
analyzer containing internal microcomputer |
 |
Virtually non-existant zero and
span drift through use of Dynamic Zero Stabilization technique invented by Dasibi |
 |
Automatic temperature and
pressure correction |
 |
Alphanumeric display |
 |
Internal diagnostic messages
available via display and optionally via RS 232 or isolated current/voltage outputs |
 |
Common electronics with Dasibi
Model 2108 NOX Analyzer, Model 3008 CO Analyzer, Model 1108 Ozone Analyzer and
Model 5009-CP Multi-Gas Calibrator |
 |
No quenching from water vapour |
 |
Built-in hydrocarbon scrubber |
 |
Standard dual analog outputs
(voltage range selectable) |
 |
Standard inlet particulate
removal system |
 |
Optional internal zero/span
check solenoid valves, microcomputer-controlled, with manual or remote activation |
Technical Specifications
| Standard Range:
|
0-2000 ppb |
| Standard
Outputs: |
0-1 ppm = 0-1 Volt |
| EPA Approved
Ranges: |
0-100 ppb, 0-200ppb, 0-550 ppb,
0-1000 ppb |
| Other Ranges: |
0-10, 20, 25, 40 and 50 ppm |
| Other "H2S"
Ranges: |
0-5 and 10 ppm |
| Analog Output
1: |
Adjustable 0-10 VFS |
| Analog Output
2: |
Adjustable 0-10 VFS |
| Noise
(Zero):* |
Less than 0.5 ppb (Fast Response
Time) |
| Less than 0.3 ppb (Slow Response
Time) |
| Lower
Detectable Limit: |
Less than 1 ppb (Fast Response
Time) |
| Less than 0.6 ppb (Slow Response
Time) |
| Zero Drift:* |
|
With Autozero: |
0 |
Without Autozero: |
Less than 1 ppb/24 hours |
| Span Drift:* |
Less than ±
0.5% per week |
| Lag Time: |
8 seconds |
| Rise
Time (0-95%):* |
90 seconds (Fast Response Time) |
| 120 seconds (Slow Response Time) |
| Fall
Time (0-95%):* |
90 seconds (Fast Response Time) |
| 120 seconds (Slow Response Time) |
Total
Interference Equivalent:
(EPA Levels) |
Less than 9 ppb |
| Precision: |
1% of reading or 1 ppb |
| Linearity: |
1% of range |
| Temperature
Range: |
20-30°C (EPA Approved) |
| 10-40°C |
| Flow Rate: |
0.5 l/min |
| Power:
|
105-125 VAC, 50-60 Hz |
| 220-240 VAC, 50-60 Hz |
Permeation Oven Subsystem Performance Specifications for
Models 4108 w/PERM & 4108 w/P&H2S
| Standard
Output: |
420-480 ppm |
| Stability: |
±
0.002 ppm |
| Repeatability: |
± 1% |
Series 4109 H2S to SO2 Convertor Specifications
| Flow Rate: |
0.5 l/min |
| Efficiency: |
96-100% (100-2000 ppb) |
| Operating
Temperature: |
700°C (approx.) |
| Power:
|
105-125 VAC, 50/60 Hz |
| 220-240 VAC, 50/60 Hz |
|